ASSEMBLY TIPS + TEST CRITERIA AND CONTROLS + POSSIBLE COMPLAINTS ASSEMBLY TIPS � Adjusting knob (11) ; Before assembling the adjusting knob (11), set the cam of the adjusting unit 10 to the indicated position. � Spring: Before replacing
3 2 4
1
0
5
2 mm
� Adjusting the hand-strap
1
1
2
N.B. � To avoid dust from getting into the pump, the suction opening has been closed with a plastic cap. Remove the plastic cap before assembling. � After the product has been repaired, it should function properly and has to meet the safety requirements as laid down and officially established at this moment.
2
3
L
M
4
S
TEST CRITERIA AND CONTROLS Voltage Total product Adapter: Primary voltage Secondary voltage Pump Motor rollers 13.0 V 220-240 V 13.0 V 400 mA if < 12.5 V adapter is defective 13.0 V 13.0 V 180 mA ± 10% 180 mA ± 10 % Ampere 380 mA ± 10 %
POSSIBLE COMPLAINTS Status 1. Switch in �OFF� position 2. Switch set to level 1 Failure Pump functions Pump doesn�t function Cause ON/OFF switch defective Adapter defective ON/OFF switch defective Internal wiring broken Defective PCB Motor pump defective Filter is clogged Electronic switch defective Defective PCB Electronic switch defective Defective PCB Sleeves got loose Vacuum stop defective Gearing motor defective Pump clogged � filter missing Internal wiring broken
3. Switch set to level 1-5 and product not placed on the skin 4. Switch set to level 1-5 and product properly placed
LED burns
Product doesn�t function
To check, if the product functions properly. - Remove sieve (13). - Set the switch to level 1. � The rollers should rotate continuously and the LED doesn�t burn. - Cover the suction opening with the thumb. � The LED lights up. - Press the vacuum stop-buttons. � The LED goes out and the rollers stop rotating.
PCS 101 222
|
|
|
> |
|